Product description
The DT1102 ROOT (PS) provides galvanic isolation and signal conversion between the transmitter (or another signal source) and the signal processing unit.
The transfer characteristics is square root. See figure above for 4-20 mA input / output.
The galvanic isolation is 3-way: the input, output and power supply are galvanically isolated of each other.
The continuous isolation voltage is 250 VACeff in all three directions
The instrument provides 20 VDC / 24 mA on the input for supplying passive two wire transmitters.
The input and output signal ranges can be selected in any combination using the front switches:
- 0-20 mA / 4-20 mA / 0-10 V / 2-10 V input
- 0-20 mA / 4-20 mA / 0-10 V / 2-10 V output
The instrument has two inputs:
- current: 0 (4) mA – 20 mA*
- voltage: 0 (2 ) V – 10 V*
The instrument has two outputs:
- current: 0 (4) mA – 20 mA*
- voltage: 0 (2 ) V – 10 V*
* it may be use only one input and one output at a time.
The device can be ordered with two types of power supply:
- DT1102 INV: 24 V ± 10% DC
- DT1102 INV PS: 230 V ± 10% AC/DC
The transfer characteristics are outstandingly good: nonlinearity < 0,02% TYP, temperature coefficient: 25 ppm / °C TYP
Front LED indicators provide information on the presence of power and possible fault conditions.