The connection terminals of the supply voltages are isolated from each other, the isolation is in compliance with the standard EN 61010-1, taking into consideration the following: |
Pollution level: | 2 |
Measurement category: | II |
Intrinsical safety data: |
Certification: | BKI 15 ATEX 0028 X |
Protection marking: | Ex II 1G Ex ia IIC T4 Ga (-20 °C < Ta < +60 °C) |
Input safety data (Pt 100): | Uo=6.6 V, Io=9 mA, Po=15 mW, Co=2 uF, Lo=10 mH |
Input safety data (4-20 mA): | Ui=30 V, Ii=100 mA, Pi=0.75 W, Ci=48 nF, Li~0 |
Limit outputs safety data: | Ui=10 V, Ii=16 mA, Pi=40 mW, Ci~0, Li~0, Uo=6.6 V, Io=0.2 mA,
Po=0.2 mW |
Input parameters: |
Input signal: | resistance of Pt100 sensor |
Connection: | 3 / 4 wire |
Measurement range: | -200 °C - +600 °C |
Measurement current: | 800 uA |
Sampling: | 15 measurement / sec. |
Number of averaged samples: | 1; 2; 4; 8; 16; 32 (selectable) |
Accuracy (refer to the nominal resistance): | < 0.3 °C @ -200°C < Tp < +400 °C
< 0.4 °C @ 400°C < Tp < 600 °C |
Output parameters: |
Analog output: |
Output: | DC current |
Leakage Current: | I < 0.1 mA @ 9 V |
Voltage drop: | U < 1 V @ 10 mA |
Load rating: | 10 V, 16 mA max. (Ex safety data) |
Refreshing time: | same as the display refreshing time |
Hysteresis: | 0 - ±99.9 |
Display: |
Display: | 4 1/2 digit, 7 segments, decimal point and negative sign |
Display unit: | LCD, 20.5 mm character height |
Display refreshing time: | 0.1; 0.3; 0.5; 1.0; 2.0 sec. (selectable) |
Manual controls: |
Manual controls: | 3 membrane push-buttons on the front cover |
Power supply: |
Power supply: | 12-30 V; protection against reverse polarity |
Ambient conditions: |
Operating temperature range: | -20 - +60 °C |
Relative humidity: | 90% (max., non-condensing) |
Place of installation: | potentially explosive area, zone 1, zone 2, safe area |
Electromagnetic compatibility (EMC)
accordance with the standard EN 61326-1 |
Immunity: | industrial area |
Noise emission: | Group 1, Class B |
Data in general: |
Housing: | moulded polycarbonate case, installable as a field or panel-instrument |
Connection: | IP 65 cable gland,
push-in direct connection |
Connection cable: | 0.25-1.5 mm2 (max.) |
Dimensions: | 167 × 87 × 90 mm
(width × height × depth) |
Weight: | 0,8 kg |
Protection: | IP 65 |
|
Detailed information see in operating instructions. The Manufacturer maintains the right to change the technical data! |