Product features
- Resistance / potentiometer inputs
- 3 or 4 wire measurement
- 0-10000 ohm measurement ranges
- Two wire 4-20 mA transmitter
- 0.05% accuracy
- TS-35 rail mounting, 12.5 mm width

The DT1005 I4 Resistance / Potentiometer Transmitter provide a 4-20 mA signal in proportional to the resistance / potentiometer connected to their input. The device is loop-powered from 4-20 mA signal.
The resistance / potentiometer can be connect in 3 or 4 wire measuring mode.
The input ranges are fix configurated, see on the above table. Different ranges for customer request also available.
Input and output is not isolated.
| Safety data: | |
|---|---|
| The connection terminals of the supply voltages are isolated from each other, the isolation is in compliance with the standard EN 61010-1, taking into consideration the following: | |
| Pollution level: | 2 |
| Measurement category: | II |
| Input parameters: | |
| Input signal: | resistance / potentiometer |
| Measurement range: | see the type designation |
| Connection: | 3 or 4 wire |
| Measuring current: | depends on the measuring range |
| Wire resistance: | 100 ohm (max) |
| Output parameters: | |
| Output signal: | 4-20 mA |
| Overloading: | 22 mA |
| Error: | < 0.05% + 10 μA (Tk: 23 °C ±2 °C) < 0.25% + 50 μA (Tk: 0 °C - 60 °C) |
| Supply-voltage effect: | practically zero |
| Power supply: | |
| Supply voltage: | 10-30 VDC loop-powered |
| Ambient conditions: | |
| Operating temperature range: | 0-60 °C (-20 - +60 °C for customer request) |
| Relative humidity: | 90% (max., non-condensing) |
| Place of installation: | cabinet |
| Electromagnetic compatibility (EMC): accordance with the standard EN 61326 |
|
| Immunity: | industrial area |
| Noise emission: | Group 1, Class B |
| General data: | |
| Housing: | terminal assembly box, rail mounting on TS-35 rail, material: polyamide PA6.6 |
| Connection: | screw terminal |
| Connection cable: | 0.25-2.5 mm2 (max.) |
| Dimensions: | 12.5 × 99 × 115 mm (width × height × depth) |
| Weight: | 0.15 kg |
| Protection: | IP 20 |
| Detailed information see in operating instructions. The Manufacturer maintains the right to change the technical data! | |
| Leaflet of product |